ChemGARD®
and ChemGARD®-RI (Radioisotope) High Performance
Fume Hoods Feature Continuous Flow Bypass for Improved
Performance. |
| Baker
ChemGARD® high performance fume hoods have
been developed by experts in containment technology.
Both ChemGARD® fume hoods employ unique
airflow characteristics within a streamlined, unobstructed
work area to provide uniform, energy-saving operation,
minimizing backdrafts, turbulence and eddy currents. |
ChemGARD®
Model FH5C
|
The
Baker ChemGARD® is the Industry’s Only
Fume Hood Which Carries the UL Classification. |
| Features |
| ChemGARD®
fume hoods are designed for unparalleled safety for
critical laboratory applications.
| |
The
continuous flow bypass design substantially
reduces roll effect, minimizes lingering
concentrations above and behind the viewscreen,
and reduces contaminant concentrations near
the edge of the sash, reducing potential exposure
hazard to personnel. |
| |
Combination
horizontal/vertical sash maintains optimum face
velocity while reducing the amount of exhausted
conditioned air to approximately one-half the
air of conventional cabinets with vertical-only
sashes, resulting in significant energy savings.
|
| |
Angled
side-entrance, airfoil and radius work surface
deliver undisturbed airflow to the work area
with minimal turbulence. |
| |
The
Airwatch II airflow monitor can be custom configured
to meet a variety of laboratory safety requirements.
|
| |
A
mass airflow sensor monitors either face velocity
or exhaust duct velocity, and activates an alarm
if monitored flow drops below safe limits. |
| |
ChemGARD®-RI
includes structural and fabrication enhancements
developed specifically for use with radioisotopes..
|
| |
Select
from three optional base cabinets to suit any
laboratory application. |
| |
All
ChemGARD® electrical and plumbing
services are fully installed and ready for easy
connection for simple, cost effective installation
and maintenance. |
|
| General
Construction Features |
| The
aerodynamic ChemGARD® design employs fixed
air slots at maximum safety configuration to alleviate
the need for cumbersome operator adjustments.
|
ChemGARD®
construction materials are fireproof.
|
|
Top-suspended,
12” wide safety plate glass sash panels are
easily removed for cleaning. |
|
Unique
internal baffle location minimizes retained
vapors in work area. |
|
ChemGARD®
fits through standard doorways and, once installed,
requires only a 2” aisle overhang. |
|
Radius
bottom airfoil minimizes turbulence. |
| |
| ChemGARD®
General Purpose Models |
|
Stainless
steel bolted, 16-gauge, Type 304 stainless steel
liner provides long-term durability, easy cleaning
and resistance to corrosion (can be epoxy coated
as an option). |
|
A
molded, dished epoxy resin work surface confines
spills. |
|
Recessed
work surface encourages proper placement of
potentially hazardous chemicals. |
|
The
continuous flow bypass design substantially reduces
roll effect, minimizes lingering concentrations above
and
behind the upper viewscreen, and virtually eliminates
contaminant concentrations near the edge of the sash,
reducing potential exposure hazard to personnel.
|
| ChemGARD®-RI
Radioisotope Models |
ChemGARD®
has been extensively tested in The Baker Company
AirFlow Management Laboratory for performance assurance.
|
|
Formed
radii, precision-engineered, 16-gauge, Type
304 stainless steel, coved corners, water-tight
welded and polished seams. |
|
Coved
corner stainless steel work surface, ¾” deep,
continuously welded to interior. |
|
Reinforced
work surface accommodates lead shielding, supports
minimum 200 lbs/sq.ft., total work surface weight
of 1,000 lbs. |
|
Stainless
steel provides long-term durability, easy cleaning
and resistance to corrosion. |
|
| Airwatch
II Airflow Monitor and Controls for Customized Operation |
The
microprocessor based Airwatch II airflow monitor is
fully programmable, offering a range of features for
safety, convenience and easier integration into the
overall laboratory safety plan.
|
The
Airwatch II airflow monitor activates an audible
alarm if operating conditions deviate from preset
parameters. |
|
The
programmable display can be set for analog and
digital, or analog only readout. |
| |
Airflow
can be measured in feet per minute (FPM) or
cubic feet per minute (CFM). |
| |
The
alarm probe monitors either face velocity or
exhaust duct velocity; the face velocity probe
is located at the front of the cabinet for easy
access and improved accuracy with horizontal
sash positions. |
|
Lock-out
options for alarm setpoints and calibration eliminate
unauthorized disabling of alarm system. |
|
Integrated
controls include flush membrane switches which
are easy to operate and easy to clean. |
|
Independent
light and duplex electrical outlet switches are
included on the main control panel. |
|
Remote
alarm contacts allow integration with existing
laboratory monitoring and safety systems. |
|
| Horizontal/Vertical
Sash Offers Advantages |
The
ChemGARD® incorporates a combination horizontal/vertical
sash designed for energy savings, operator safety and
convenience.
|
Combination
horizontal/vertical sash provides a protective
physical barrier with full vertical access to
the work area when using either horizontal or
vertical configuration. |
|
Full
vertical access allows easy installation of
experimental apparatus. |
|
Top
suspended 12” wide safety plate glass panels
are easily removed for cleaning. |
|
Side
panels are flush with bell mouth inlet for streamlined
airflow; the radiused bottom airfoil minimizes
turbulence. |
|
Sliding
glass panels can be positioned for increased operator
safety. |
|
Ball
bearing, top hung horizontal sash support provides
smoother sash operation. |
|
Hood
face opening is reduced 50% for energy savings.
|
|
| Lighting
and Electrical System |
|
A
sealed, fluorescent light fixture is located above
gasketed safety-plate glass; fluorescent lamps
are replaced from the front. |
|
Solid-state
lamp ballasts reduce heat output and improve
efficiency. |
|
Two
GFI duplex outlets are protected by a single
10-amp self-resetting circuit breaker to protect
personnel and minimize tripping of breakers.
|
|
ChemGARD®
is the industry’s only fume hood which carries
the UL Classification for electrical and personal
protection. |
|
| Utilities |
| |
Color-coded service fixtures for air, gas, vacuum,
cold or DI water and steam, include convenient
remote control valves. |
|
The
cabinet includes one fixture and one petcock
standard, plus seven plugged penetrations for
additional optional fixtures. |
|
Pre-studded
exhaust outlet, 10” diameter, accepts flanged
exhaust connection. |
|
Airwatch
II airflow monitor. |
|
| ChemGARD®
Design Features |
|
Combination
horizontal/vertical sash enhances both operator
safety and energy savings. |
|
Sight-tight
continuous bypass duct minimizes contaminant
concentrations near the edge of the sash, reducing
potential exposure hazard to personnel. |
|
Airwatch
II airflow monitor provides for customized alarm/monitor
configurations. |
|
Angled
entrance, airfoil and radius work surface for
undisturbed airflow into work area. |
|
Color-coded
service fixtures with convenient remote control
valves, positioned outside the air stream. |
|
Counterbalanced
vertical sash for smooth, single-handed operation |
|
Recessed
work surface confines spills. |
|
All
stainless steel construction for durability and
easy cleaning. |
|
Optional
general purpose, solvent, or acid base cabinet
provides storage area. |
| |
|
|
| Options
and Accessories |
For
convenience, most options, accessories and modifications
are factory installed and should be specified when
ordering. Commonly requested options are listed below.
For detailed information on accessories and modifications,
contact The Baker Company.
|
Additional
service fixtures (air, gas, vacuum, cold or DI
water, steam) |
|
Cold
water gooseneck faucet |
|
Distilled/DI
water gooseneck |
|
Vacuum
breakers |
|
Aspirator |
|
Epoxy
resin sink |
|
Base
cabinet (see below) |
|
Stainless
steel peg board |
|
Tubular
base stand |
|
Vertical
only sash |
|
Cupsink |
|
Stainless
steel work surface |
|
| Base
Cabinets |
| Choose
From Three Cabinet Bases, Four Base Sizes. |
| General
Purpose Base Cabinet |
The
general purpose base cabinet has removable front and
side panels, plus a 10” deep chase at the rear to
allow room for services. Other features include:
|
Reinforced
shelf, adjustable on 1” centers. |
|
Dual
panel doors, 1½” thick, secured with 3” wide
continuous hinge, 3-point latch. |
|
18-gauge,
cold-rolled steel bottom pan, recessed 1½”,
liquid tight. |
|
2½”
riser with 3/8” leveling feet. |
|
16-gauge,
cold-rolled steel side panels. |
|
Epoxy
powder coated throughout. |
|
| Acid
Storage Base Cabinet |
The
acid storage base cabinet has an acid-resistant polypropylene
drip tray, shelf liner and (optional) vent assembly
with flexible polypropylene hose designed to “T” into
the fume hood exhaust plenum.
|
2”
diameter plugged penetrations in back wall, and
removable side and rear panels with a 10” deep
chase at the rear to allow for installation of
the optional vent package. |
|
The
vent package offers from 10 to 15 air changes
per hour, depending on cabinet size. |
|
Reinforced
steel shelf, adjustable on 1” centers. |
|
Dual
panel doors, 1½” thick, secured with 3” wide continuous
hinge, 3-point latch. |
|
18-gauge,
cold-rolled steel bottom pan, recessed 1½”, liquid
tight. |
|
2½”
riser with 3/8” leveling feet. |
|
16-gauge,
cold-rolled steel side panels. |
|
Epoxy
powder coated throughout. |
|
| Solvent
Storage Base Cabinet |
The
solvent storage base cabinet meets NFPA 30 and UL
1275 criteria for safety and performance.
|
Double wall construction creates a minimum 1½”
air gap throughout the cabinet |
|
Interior
back and side panels are 18-gauge stainless
steel monolithic construction welded to top
and bottom panels to form a unitized inner cabinet. |
|
The
18-gauge stainless steel bottom pan is recessed
2” and welded liquid tight. |
|
2”
diameter plugged penetrations in back wall, and
two front removable rear panels provide access
to the 10” deep chase at the rear to allow for
installation of the (optional) vent package. |
|
Access
panels are located at each end of the chase. |
|
An
optional vent package offers from 10 to 15 air
changes per hour, depending on cabinet size; the
vent assembly with flexible hose is designed to
“T” into the fume hood exhaust plenum. |
|
A
reinforced T304 stainless steel shelf is adjustable
on 1” centers. |
|
Dual
panel doors, 1½” thick, with 3” wide continuous
hinge, and secured with a 3-point latch; optional
louver doors are available. |
|
2½”
riser with 3/8” leveling feet. |
|
16-gauge,
cold-rolled steel side panels, epoxy powder coated. |
|
| Base
Cabinet Description and Model Numbers
|
| General
Purpose |
epoxy
resin |
FHB3GP
|
FHB4GP
|
FHB5GP |
FHB6GP |
| Acid
Storage |
polypropylene |
FHB3ACID |
FHB4ACID |
FHB5ACID |
FHB6ACID |
| Solvent
Storage |
stainless
steel |
FHB3SOL |
FHB4SOL |
FHB5SOL |
FHB6SOL |
|
|
To Order |
|
Caution:
The
adequacy of this containment cabinet for
the user’s personal safety, as with any
containment cabinet, should be determined
by an industrial hygienist or safety officer.
Site preparation information, architectural
drawings, detailed dimensions and purchase
specifications are available.
|
|
| Summary
Specification and Site Preparation
|
ChemGARD®
General Purpose,
High Performance |
ChemGARD®-4 |
ChemGARD®-5 |
ChemGARD®-6 |
ChemGARD®-8
|
ChemGARD®
Radioisotope
Model, High Performance |
ChemGARD®-RI4 |
ChemGARD®-RI5 |
ChemGARD®-RI6 |
n/a |
Overall
Dimensions
(Nominal) |
48”W
x 33¾”
F-B x 59”H |
60”W
x 33¾”
F-B x 59”H |
72”W
x 33¾”
F-B x 59”H |
96”W
x 33¾”
F-B x 59”H |
Optional
Base Cabinet
Dimensions |
48”W
x 29”F-B
x 35¾”H(min.) |
60”W
x 29”F-B
x 35¾”H(min.) |
72”W
x 29”F-B
x 35¾”H(min.) |
96”W
x 29”F-B
x 35¾”H(min.) |
Vertical
|
40½”W
x 27”H
|
52½”W
x 27”H |
64½”W
x 27”H |
88½”W
x 27”H |
| Horizontal |
16¼”W
x 253/8”H |
16¼”W
x 253/8”H |
28¼”W
x 253/8”H |
40¼”W
x 253/8”H |
| Vertical |
40½”W
x 28½”H |
52½”W
x 28½”H |
64½”W
x 28½”H |
88½”W
x 28½”H |
100
FPM with vertical
sash at maximum height |
808
CFM @0.42
in H2O |
1047
CFM @0.70
in H2O |
1286
CFM @0.85
in H2O |
1754
CFM @0.52
in H2O |
100
FPM with horizontal
sash at maximum width |
374
CFM @0.11
in H2O |
405
CFM @0.11
in H2O |
644
CFM @0.25
in H2O |
900
CFM @0.16
in H2O |
100
FPM with vertical
sash at maximum height |
808
CFM @0.42
in H2O |
1047
CFM @0.70
in H2O |
1286
CFM @0.85
in H2O |
1754
CFM @0.52
in H2O |
| Weight |
400
lbs |
500
lbs |
600
lbs |
800
lbs |
| Illumination
(Foot-Candles) |
50
|
50 |
60 |
50 |
| Electrical |
115V,
60Hz |
115V,
60Hz |
115V,
60Hz |
115V,
60Hz |
| Total
Running Amps |
11.1
|
11.1
|
11.2
|
12.1 |
| GFI
Duplex Outlets |
2
w/10 amp
breaker |
2
w/10 amp
breaker |
2
w/10 amp
breaker |
2
w/10 amp
breaker |
| Auxiliary
Switch Rating |
20
amps
@115V, AC |
20
amps
@115V, AC |
20
amps
@115V, AC |
20
amps
@115V, AC |
| |
10
amps
@250V, AC |
10
amps
@250V, AC |
10
amps
@250V, AC |
10
amps
@250V, AC |
| Plumbing |
3/8”
female
brass fittings |
3/8”
female
brass fittings |
3/8”
female
brass fittings |
3/8”
female
brass fittings |
|
| |
|
| Purchase
Specifications |
| ChemGARD®-RI |
|
Every radioisotope fume hood design shall have a certified
copy of the ANSI/ASHRAE-110 Fume Hood Performance
Test available at the request of the purchaser.
|
|
The fume hood interior shall be constructed of 16-gauge,
Type 304 stainless steel. |
|
The front columns shall have an ideal 40°angle with an
aerodynamic shape on the airfoil, bottom edge
of sash and the front edge of the 1” thick stainless
steel work surface. |
|
Rear baffle shall be front removable with fixed rear exhaust
slots. |
|
The
side wall and front columns shall be seamless,
one-piece construction. The continuous bypass
inlet shall be sight-tight. |
|
Service
fixture(s) shall be color coded with chemical
resistant coating and have a remote control valve. |
|
The
sash shall be a combination horizontal/vertical
type with a vertical sash as an option. |
|
The
light fixture shall be front accessible, located
external to the hood, and sealed from the interior
by gasketed safety plate glass. |
|
The fume hood shall be pre-wired and pre-plumbed unless
otherwise specified by the purchaser. |
|
Each
fume hood shall be equipped with the Airwatch
II airflow monitor and alarm system. |
|
Every
fume hood shall have a front removable continuous
bypass duct to reduce the roll effect. |
|
The
work surface shall be water-tight, radiused, ¾”
dished, stainless steel. |
|
The
interior shall contain no unwelded or unground
seams. |
|
The
interior work area (i.e. work surface, side, rear,
and top walls) shall be monolithic construction,
with all continuous welds, ground smooth, polished,
with coved corners and radiused seams to provide
a continuous surface. |
|
The
exhaust outlet shall be pre-studded to accept
a flanged exhaust duct. |
|
The
work surface shall be capable of supporting 200
lbs/sq. ft. minimum, up to 1,000 lbs per fume
hood. |
|
| Warranty |
The
Baker Company, Inc., expressly represents and warrants
all goods (a) to be as specified (and described) in
The Baker Company catalogs and literature, and (b)
to be free under normal use, service and testing (all
as described in The Baker Company catalogs and literature)
from defects in material and workmanship for a period
of twelve months from the invoice date.
The
exclusive remedy for any breach or violation of this
warranty is as follows: The Baker Company, Inc., will
F.O.B. Sanford, Maine, furnish without charge repairs
to or replacement of the parts or equipment which
proved defective in material or workmanship. No claim
may be made for any incidental or consequential damages.
THIS
WARRANTY IS EXPRESSLY IN LIEU OF ALL OTHER WARRANTIES,
EXPRESSED OR IMPLIED, INCLUDING ANY IMPLIED WARRANTY
OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE
UNLESS OTHERWISE AGREED IN WRITING SIGNED BY THE BAKER
COMPANY. (THE BAKER COMPANY SHALL NOT BE RESPONSIBLE
FOR ANY IMPROPER USE, INSTALLATION, SERVICE OR TESTING
OF THE GOODS. |
|
|
|