Our Products
 
Animal Housing Solution
 
Shelf
 
Individually Ventilated Cages
   
  • SmartRack
  • RowZone
  • IPG System
  •   Animal Health Monitoring - CRL
      Immunocomb
      Serology Reagent
    Animal Identification System
      Surgery Platform
      Restrainers for Tail Injection
    Aquatic Research Solutions
     
    Standalone
     
    Benchtop
     
    Multirack
     
    Frog Rack
    Bio Decontamination System
      Why Need of VHP
      Chemdi® VHP Chemical Indicator
      VHP® 1000ED Biodecontamination System
      VHP® 100P Biodecontamination System
      Vaprox® Hydrogen Peroxide Sterilant
    Aseptic and Container Isolators
      Aseptic Isolators
      Containment Isolators
    Water Purity Systems
      PSG
      MEWS
      High Purity Water System
      Finn-Aqua® T-Series Pure Steam
    Sterilization Solution
      Sterilization-Res/GMP
      CED
      Sterilizers
       
  • Amsco® Century® Small Steam Sterilizer
  • Amsco® Century® Medium Steam Sterilizers 26 x 26
  • Amsco® Century® Medium Steam Sterilizers 26 x 37.5
  • Finn-Aqua® Steam Sterilizer
  • Finn-Aqua® BIO Pharma Series(BPS) GMP Steam Sterilizers
  • Finn-Aqua® Life Science Research (LSR) Steam Sterilizer
  • Washing Solutions
      Research /GMP
       
      Animal
       
    Containment
      Animal Products
       
      Lab Products
       
    Cleaning Chemistries
    Toxicology & Behavioural Monitoring
     
    Behavior
    Drinking & Feeding
    Calorimetry
    LabMaster
    PhenoMaster
    Analgesia
    Physiology
    Respiration & Anesthesia
    Inhalation
    Stereotaxic Instruments
    Microdialysis
    Isolated Organs
    Microtomes
    Pumps & Infusion
    Surgery & Handling
     
     
     
    Other Products
      Barrier Isolator
      Automated Washing System
       
      Cage Cleaners
      Bottle Filler
     
    Fume Hoods
    ChemGARD® and ChemGARD®-RI (Radioisotope) High Performance Fume Hoods Feature Continuous Flow Bypass for Improved Performance.

    Baker ChemGARD® high performance fume hoods have been developed by experts in containment technology. Both ChemGARD® fume hoods employ unique airflow characteristics within a streamlined, unobstructed work area to provide uniform, energy-saving operation, minimizing backdrafts, turbulence and eddy currents.


    ChemGARD® Model FH5C
    The Baker ChemGARD® is the Industry’s Only Fume Hood Which Carries the UL Classification.
    Features
    ChemGARD® fume hoods are designed for unparalleled safety for critical laboratory applications.
    The continuous flow bypass design substantially reduces roll effect, minimizes lingering concentrations above and behind the viewscreen, and reduces contaminant concentrations near the edge of the sash, reducing potential exposure hazard to personnel.
    Combination horizontal/vertical sash maintains optimum face velocity while reducing the amount of exhausted conditioned air to approximately one-half the air of conventional cabinets with vertical-only sashes, resulting in significant energy savings.
    Angled side-entrance, airfoil and radius work surface deliver undisturbed airflow to the work area with minimal turbulence.
    The Airwatch II airflow monitor can be custom configured to meet a variety of laboratory safety requirements.
    A mass airflow sensor monitors either face velocity or exhaust duct velocity, and activates an alarm if monitored flow drops below safe limits.
    ChemGARD®-RI includes structural and fabrication enhancements developed specifically for use with radioisotopes..
    Select from three optional base cabinets to suit any laboratory application.
    All ChemGARD® electrical and plumbing services are fully installed and ready for easy connection for simple, cost effective installation and maintenance.
    General Construction Features
    The aerodynamic ChemGARD® design employs fixed air slots at maximum safety configuration to alleviate the need for cumbersome operator adjustments.
    ChemGARD® construction materials are fireproof.
    Top-suspended, 12” wide safety plate glass sash panels are easily removed for cleaning.
    Unique internal baffle location minimizes retained vapors in work area.
    ChemGARD® fits through standard doorways and, once installed, requires only a 2” aisle overhang.
    Radius bottom airfoil minimizes turbulence.
     
    ChemGARD® General Purpose Models
    Stainless steel bolted, 16-gauge, Type 304 stainless steel liner provides long-term durability, easy cleaning and resistance to corrosion (can be epoxy coated as an option).
    A molded, dished epoxy resin work surface confines spills.
    Recessed work surface encourages proper placement of potentially hazardous chemicals.

    The continuous flow bypass design substantially reduces roll effect, minimizes lingering concentrations above and
    behind the upper viewscreen, and virtually eliminates contaminant concentrations near the edge of the sash, reducing potential exposure hazard to personnel
    .
    ChemGARD®-RI Radioisotope Models

    ChemGARD® has been extensively tested in The Baker Company
    AirFlow Management Laboratory for performance assurance.

    Formed radii, precision-engineered, 16-gauge, Type 304 stainless steel, coved corners, water-tight welded and polished seams.
    Coved corner stainless steel work surface, ¾” deep, continuously welded to interior.
    Reinforced work surface accommodates lead shielding, supports minimum 200 lbs/sq.ft., total work surface weight of 1,000 lbs.
    Stainless steel provides long-term durability, easy cleaning and resistance to corrosion.
    Airwatch II Airflow Monitor and Controls for Customized Operation
    The microprocessor based Airwatch II airflow monitor is fully programmable, offering a range of features for safety, convenience and easier integration into the overall laboratory safety plan.

    The Airwatch II airflow monitor activates an audible alarm if operating conditions deviate from preset parameters.
    The programmable display can be set for analog and digital, or analog only readout.
    Airflow can be measured in feet per minute (FPM) or cubic feet per minute (CFM).
    The alarm probe monitors either face velocity or exhaust duct velocity; the face velocity probe is located at the front of the cabinet for easy access and improved accuracy with horizontal sash positions.
    Lock-out options for alarm setpoints and calibration eliminate unauthorized disabling of alarm system.
    Integrated controls include flush membrane switches which are easy to operate and easy to clean.
    Independent light and duplex electrical outlet switches are included on the main control panel.
    Remote alarm contacts allow integration with existing laboratory monitoring and safety systems.
    Horizontal/Vertical Sash Offers Advantages
    The ChemGARD® incorporates a combination horizontal/vertical sash designed for energy savings, operator safety and convenience.

    Combination horizontal/vertical sash provides a protective physical barrier with full vertical access to the work area when using either horizontal or vertical configuration.
    Full vertical access allows easy installation of experimental apparatus.
    Top suspended 12” wide safety plate glass panels are easily removed for cleaning.
    Side panels are flush with bell mouth inlet for streamlined airflow; the radiused bottom airfoil minimizes turbulence.
    Sliding glass panels can be positioned for increased operator safety.
    Ball bearing, top hung horizontal sash support provides smoother sash operation.
    Hood face opening is reduced 50% for energy savings.
    Lighting and Electrical System
    A sealed, fluorescent light fixture is located above gasketed safety-plate glass; fluorescent lamps are replaced from the front.
    Solid-state lamp ballasts reduce heat output and improve efficiency.
    Two GFI duplex outlets are protected by a single 10-amp self-resetting circuit breaker to protect personnel and minimize tripping of breakers.
    ChemGARD® is the industry’s only fume hood which carries the UL Classification for electrical and personal protection.
    Utilities
    Color-coded service fixtures for air, gas, vacuum, cold or DI water and steam, include convenient remote control valves.
    The cabinet includes one fixture and one petcock standard, plus seven plugged penetrations for additional optional fixtures.
    Pre-studded exhaust outlet, 10” diameter, accepts flanged exhaust connection.
    Airwatch II airflow monitor.
    ChemGARD® Design Features
    Combination horizontal/vertical sash enhances both operator safety and energy savings.
    Sight-tight continuous bypass duct minimizes contaminant concentrations near the edge of the sash, reducing potential exposure hazard to personnel.
    Airwatch II airflow monitor provides for customized alarm/monitor configurations.
    Angled entrance, airfoil and radius work surface for undisturbed airflow into work area.
    Color-coded service fixtures with convenient remote control valves, positioned outside the air stream.
    Counterbalanced vertical sash for smooth, single-handed operation
    Recessed work surface confines spills.
    All stainless steel construction for durability and easy cleaning.
    Optional general purpose, solvent, or acid base cabinet provides storage area.
     

    Options and Accessories
    For convenience, most options, accessories and modifications are factory installed and should be specified when ordering. Commonly requested options are listed below. For detailed information on accessories and modifications, contact The Baker Company.

    Additional service fixtures (air, gas, vacuum, cold or DI water, steam)
    Cold water gooseneck faucet
    Distilled/DI water gooseneck
    Vacuum breakers
    Aspirator
    Epoxy resin sink
    Base cabinet (see below)
    Stainless steel peg board
    Tubular base stand
    Vertical only sash
    Cupsink
    Stainless steel work surface
    Base Cabinets
    Choose From Three Cabinet Bases, Four Base Sizes.
    General Purpose Base Cabinet
    The general purpose base cabinet has removable front and side panels, plus a 10” deep chase at the rear to allow room for services. Other features include:

    Reinforced shelf, adjustable on 1” centers.
    Dual panel doors, 1½” thick, secured with 3” wide continuous hinge, 3-point latch.
    18-gauge, cold-rolled steel bottom pan, recessed 1½”, liquid tight.
    2½” riser with 3/8” leveling feet.
    16-gauge, cold-rolled steel side panels.
    Epoxy powder coated throughout.
    Acid Storage Base Cabinet
    The acid storage base cabinet has an acid-resistant polypropylene drip tray, shelf liner and (optional) vent assembly with flexible polypropylene hose designed to “T” into the fume hood exhaust plenum.
    2” diameter plugged penetrations in back wall, and removable side and rear panels with a 10” deep chase at the rear to allow for installation of the optional vent package.
    The vent package offers from 10 to 15 air changes per hour, depending on cabinet size.
    Reinforced steel shelf, adjustable on 1” centers.
    Dual panel doors, 1½” thick, secured with 3” wide continuous hinge, 3-point latch.
    18-gauge, cold-rolled steel bottom pan, recessed 1½”, liquid tight.
    2½” riser with 3/8” leveling feet.
    16-gauge, cold-rolled steel side panels.
    Epoxy powder coated throughout.
    Solvent Storage Base Cabinet
    The solvent storage base cabinet meets NFPA 30 and UL 1275 criteria for safety and performance.
    Double wall construction creates a minimum 1½” air gap throughout the cabinet
    Interior back and side panels are 18-gauge stainless steel monolithic construction welded to top and bottom panels to form a unitized inner cabinet.
    The 18-gauge stainless steel bottom pan is recessed 2” and welded liquid tight.
    2” diameter plugged penetrations in back wall, and two front removable rear panels provide access to the 10” deep chase at the rear to allow for installation of the (optional) vent package.
    Access panels are located at each end of the chase.
    An optional vent package offers from 10 to 15 air changes per hour, depending on cabinet size; the vent assembly with flexible hose is designed to “T” into the fume hood exhaust plenum.
    A reinforced T304 stainless steel shelf is adjustable on 1” centers.
    Dual panel doors, 1½” thick, with 3” wide continuous hinge, and secured with a 3-point latch; optional louver doors are available.
    2½” riser with 3/8” leveling feet.
    16-gauge, cold-rolled steel side panels, epoxy powder coated.
    Base Cabinet Description and Model Numbers
    Base Cabinet Surface Material 3-foot 4-foot 5-foot 6-foot
    General Purpose epoxy resin FHB3GP FHB4GP FHB5GP FHB6GP
    Acid Storage polypropylene FHB3ACID FHB4ACID FHB5ACID FHB6ACID
    Solvent Storage stainless steel FHB3SOL FHB4SOL FHB5SOL FHB6SOL
    To Order

    Caution:
    The adequacy of this containment cabinet for the user’s personal safety, as with any containment cabinet, should be determined by an industrial hygienist or safety officer. Site preparation information, architectural drawings, detailed dimensions and purchase specifications are available.

    Summary Specification and Site Preparation
     ChemGARD® and ChemGARD®-RI High Performance Fume Hood
    ChemGARD®
    General Purpose,
    High Performance
    ChemGARD®-4 ChemGARD®-5 ChemGARD®-6 ChemGARD®-8
    ChemGARD® Radioisotope
    Model, High Performance
    ChemGARD®-RI4 ChemGARD®-RI5 ChemGARD®-RI6 n/a
    Overall Dimensions
    (Nominal)
    48”W x 33¾”
    F-B x 59”H
    60”W x 33¾”
    F-B x 59”H
    72”W x 33¾”
    F-B x 59”H
    96”W x 33¾”
    F-B x 59”H
    Optional Base Cabinet
    Dimensions
    48”W x 29”F-B
    x 35¾”H(min.)
    60”W x 29”F-B
    x 35¾”H(min.)
    72”W x 29”F-B
    x 35¾”H(min.)
    96”W x 29”F-B
    x 35¾”H(min.)
     Combination Sash, Maximum Opening
    Vertical
    40½”W x 27”H
    52½”W x 27”H 64½”W x 27”H 88½”W x 27”H
    Horizontal 16¼”W
    x 253/8”H
    16¼”W
    x 253/8”H
    28¼”W
    x 253/8”H
    40¼”W
    x 253/8”H
     Vertical Sash, Maximum Opening
    Vertical 40½”W
    x 28½”H
    52½”W
    x 28½”H
    64½”W
    x 28½”H
    88½”W
    x 28½”H
     Exhaust Requirements, Combination Sash
    100 FPM with vertical
    sash at maximum height
    808 CFM @0.42
    in H2O
    1047 CFM @0.70
    in H2O
    1286 CFM @0.85
    in H2O
    1754 CFM @0.52
    in H2O
    100 FPM with horizontal
    sash at maximum width
    374 CFM @0.11
    in H2O
    405 CFM @0.11
    in H2O
    644 CFM @0.25
    in H2O
    900 CFM @0.16
    in H2O
     Exhaust Requirements, Vertical Sash
    100 FPM with vertical
    sash at maximum height
    808 CFM @0.42
    in H2O
    1047 CFM @0.70
    in H2O
    1286 CFM @0.85
    in H2O
    1754 CFM @0.52
    in H2O
    Weight 400 lbs 500 lbs 600 lbs 800 lbs
    Illumination (Foot-Candles) 50 50 60 50
    Electrical 115V, 60Hz 115V, 60Hz 115V, 60Hz 115V, 60Hz
    Total Running Amps 11.1 11.1 11.2 12.1
    GFI Duplex Outlets 2 w/10 amp
    breaker
    2 w/10 amp
    breaker
    2 w/10 amp
    breaker
    2 w/10 amp
    breaker
    Auxiliary Switch Rating 20 amps
    @115V, AC
    20 amps
    @115V, AC
    20 amps
    @115V, AC
    20 amps
    @115V, AC
      10 amps
    @250V, AC
    10 amps
    @250V, AC
    10 amps
    @250V, AC
    10 amps
    @250V, AC
    Plumbing 3/8” female
    brass fittings
    3/8” female
    brass fittings
    3/8” female
    brass fittings
    3/8” female
    brass fittings
     
    Purchase Specifications
    ChemGARD®-RI
    Every radioisotope fume hood design shall have a certified copy of the ANSI/ASHRAE-110 Fume Hood Performance Test available at the request of the purchaser.
    The fume hood interior shall be constructed of 16-gauge, Type 304 stainless steel.
    The front columns shall have an ideal 40°angle with an aerodynamic shape on the airfoil, bottom edge of sash and the front edge of the 1” thick stainless steel work surface.
    Rear baffle shall be front removable with fixed rear exhaust slots.
    The side wall and front columns shall be seamless, one-piece construction. The continuous bypass inlet shall be sight-tight.
    Service fixture(s) shall be color coded with chemical resistant coating and have a remote control valve.
    The sash shall be a combination horizontal/vertical type with a vertical sash as an option.
    The light fixture shall be front accessible, located external to the hood, and sealed from the interior by gasketed safety plate glass.
    The fume hood shall be pre-wired and pre-plumbed unless otherwise specified by the purchaser.
    Each fume hood shall be equipped with the Airwatch II airflow monitor and alarm system.
    Every fume hood shall have a front removable continuous bypass duct to reduce the roll effect.
    The work surface shall be water-tight, radiused, ¾” dished, stainless steel.
    The interior shall contain no unwelded or unground seams.
    The interior work area (i.e. work surface, side, rear, and top walls) shall be monolithic construction, with all continuous welds, ground smooth, polished, with coved corners and radiused seams to provide a continuous surface.
    The exhaust outlet shall be pre-studded to accept a flanged exhaust duct.
    The work surface shall be capable of supporting 200 lbs/sq. ft. minimum, up to 1,000 lbs per fume hood.
    Warranty

    The Baker Company, Inc., expressly represents and warrants all goods (a) to be as specified (and described) in The Baker Company catalogs and literature, and (b) to be free under normal use, service and testing (all as described in The Baker Company catalogs and literature) from defects in material and workmanship for a period of twelve months from the invoice date.

    The exclusive remedy for any breach or violation of this warranty is as follows: The Baker Company, Inc., will F.O.B. Sanford, Maine, furnish without charge repairs to or replacement of the parts or equipment which proved defective in material or workmanship. No claim may be made for any incidental or consequential damages.

    THIS WARRANTY IS EXPRESSLY IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE UNLESS OTHERWISE AGREED IN WRITING SIGNED BY THE BAKER COMPANY. (THE BAKER COMPANY SHALL NOT BE RESPONSIBLE FOR ANY IMPROPER USE, INSTALLATION, SERVICE OR TESTING OF THE GOODS.