Primary,
Under the Work Surface HEPA Filters, Supply HEPA Filter
and Direct Exhaust Airflow System Combine to Increase
Protection From Particulates, While Preventing Build-Up
of Chemical Vapors and Gases. |
Class
II, Type B1—Vertical Flow Biological Safety Cabinet
The Baker Company NCB-D® Class II, Type
B1 Biological Safety Cabinet meets or exceeds specifications
for the NCI-1 cabinet developed by the National
Cancer Institute.
The
NCB-D® is designed for biological work,
product preparation and experimentation involving
agents of low and moderate risk (Biosafety Levels
1, 2 or 3).
The cabinet exceeds design requirements of NCI Specification
“General Purpose Clean Air Biological Safety Cabinet”
(Class II, Type B Safety Cabinet dated July 6, 1976)
and National Sanitation Foundation (NSF) International
Standard #49.
Applications
for Class II, Type B1 cabinets include microbiological
studies, cell culture, pharmaceutical procedures
and toxicology.
According
to NCI, this Class II, Type B1 cabinet may be used
for work with chemical carcinogens in microgram
quantities (Laboratory Safety Monograph, NIH, January
1979).
|
NCB-D®
HEPA-filtered, particle free supply plenum plus approximately
60% direct exhaust from work area helps prevent build-up
of chemical vapors and gases.
 |
 |
 |
CSA
Listed |
Meets
NSF
Standards |
Underwriter
Laboratories
Listed |
|
| NCB-D®
Exceeds Class II, Type B1 Criteria |
|
| Features |
A high velocity suction channel prevents gases or
particles from migrating behind the viewscreen and
entering the work area.
Sliding 10,º ergonomically sloped viewscreen provides
easy access to the work area.
Mass Airflow Monitor automatically activates an
audio/visual alarm if exhaust airflow in the duct
decreases below the calibrated set point.
A
non-glare satin finish work surface with wide radius
corners improves user comfort and simplifies cleaning.
The HEPA filters located below the work surface
provide an added measure of protection by eliminating
contamination in the blower/motor area and positive
pressure areas of the cabinet.
Positive pressure plenums are downstream of HEPA
filters and are completely surrounded by negative
pressure.
Padded ergonomic arm rest.
|
| Double
HEPA Filters for Added Containment and Internal Protection |
| |
| NCB-D®
Airflow |
| |
|
|